  1. Purpose

To help EAP Sensor Evaluation Kits users

  1. Scope
  • Set up
  • Connecting to App
  • App Features

*Note: Instruction to cover both SNES-KIT-B-01 & SNES-KIT-B-01-U Evaluation Kits

  1. Parts


Description

Picture

USB battery

Micro-USB Cable

Sensor Cable

Transmitter


SNES-KIT-B-01-U

 


SNES-KIT-B-01

EAP Sensor











  1. Kit Set Up
    • Connect USB battery to BLE transmitter using the micro USB cable
      • Make sure the battery is fully charged
      • Do not connect the battery to the transmitter while at charge
      • Connecting battery to transmitter while at charge introduces noise
    • Connect the Sensor Cable to Transmitter
      • Insert sensor cable audio jack into the audio input on the BLE transmitter
    • Connect the sensor to the sensor cable
      • Sensor Pin Outs are #2 & 5, meaning that the sensor cable pins should be connected to the pin #2 & 5 on the blue connector
    • Turn on the battery
      • Battery light is blue when charged and is red when charging is needed



  1. Connecting to App
    • Download the "SensoNODE" App from Apple Store
      • Search for "SensoNODE" or "Scout Mobile"

  • Android App is not available for the time being but is expected for release in Jan 2016
    • Set up your kit and turn on the battery
    • Fire up the App
    • Connect the BLE transmitter to the SensoNODE app using the ((plus)) sign in the middle of screen or the + sign at the top-right corner
      • Hit "+" to add sensor
      • Select the sensor you like to monitor
      • Sensor identification is BLE radio serial number (See back of BLE Radio)
      • Change the name or color ID of the sensor if desired
      • Hit "Save"



  1. App Features
    • Adding a widget
      • Double Click on the Widget folder ( Square with + in the middle)
      • Select the desired widgets
      • Hit "Back" at the top-left corner. Now you see the list of selected Widgets
      • Hit desired widgets to monitor sensor out put



    • Removing a Widget
      • To remove a widget, double click on the widget icon
      • Hit "Remove"



  • Changing the read out from EAP Percentage to EAP Raw
    • To change EAP sensor read out from Percentage to Raw hit "EAP Percentage"
    • From the pop-up menu select "Show EAP Raw"
    • To see EAP Raw count select the read out icon






*Notes:

  • Selected Widgets do not roll over from Percentage Mode to Raw Mode
  • Desired widgets needs to be selected independently for each percentage and raw modes
  • In raw mode, user get to see the count numbers.








    • Calibration
      • You may notice that connected sensors might be off by up to 5% before calibration
      • This is because the app has a default "Count Number" setting that needs to be changed for each sensor to improve accuracy
      • For how to access "Count Numbers" see previous section
      • To calibrate, collect the Count numbers at 0% and 100% strain. Record/note the Counts numbers when the sensor is relaxed for 0% and stretch the sensor 50 mm and record the count number for 100%. These two numbers are min and max that needs to be replaced with App's default setting.
      • For calibration highlight the "Connection" Icon with green color
      • Hit the pro icon with letter i
      • Replace the calibration low and calibration high numbers with the numbers collected at 0% and 100% for the specific sensor you are calibrating.