To help EAP Sensor Evaluation Kits users

...

  • Set up
  • Connecting to App
  • App Features

*Note: Instruction to cover both SNES-KIT-B-01 & SNES-KIT-B-01-U Evaluation Kits.

Parts

Description

Picture

USB battery

Micro-USB Cable

Sensor Cable

Transmitter


SNES-KIT-B-01-U

 


SNES-KIT-B-01

EAP Sensor

 

Kit Set Up

  1. Connect USB battery to BLE transmitter using the micro USB cable

    • Make sure the battery is fully charged

    • Do not connect the battery to the transmitter while at charge

    • Connecting battery to transmitter while at charge introduces noise

  2. Connect the Sensor Cable to Transmitter
    • Insert sensor cable audio jack into the audio input on the BLE transmitter
  3. Connect the sensor to the sensor cable
    • Sensor Pin Outs are #2 & 5, meaning that the sensor cable pins should be connected to the pin #2 & 5 on the blue connector
  4. Turn on the battery
    • Battery light is blue when charged and is red when charging is needed


Image Removed Image Added

Connecting to App

  1. Download the "SensoNODE" App from Apple Store

    • Search for "SensoNODE" or "Scout Mobile"

      Note: Android App is not available for the time being but is expected for release in Jan 2016

  2. Set up your kit and turn on the battery

  3. Fire up the App
  4. Connect the BLE transmitter to the SensoNODE app using the ((plus)) sign in the middle of screen or the + sign at the top-right corner
    • Hit "+" to add sensor
    • Select the sensor you like to monitor
    • Sensor identification is BLE radio serial number (See back of BLE Radio)
    • Change the name or color ID of the sensor if desired
    • Hit "Save"

...

  1. Adding a widget
    • Double Click on the Widget folder ( Square with + in the middle)
    • Select the desired widgets
    • Hit "Back" at the top-left corner. Now you see the list of selected Widgets
    • Hit desired widgets to monitor sensor out put


  2. Removing a Widget
    • To remove a widget, double click on the widget icon
    • Hit "Remove"


  3. Changing the read out from EAP Percentage to EAP Raw
    • To change EAP sensor read out from Percentage to Raw hit "EAP Percentage"
    • From the pop-up menu select "Show EAP Raw"
    • To see EAP Raw count select the read out icon

    *

    Notes: 
    • Selected Widgets do not roll over from Percentage Mode to Raw Mode
    • Desired widgets needs to be selected independently for each percentage and raw modes
    • In raw mode, user get to see the count numbers.

  4. Calibration
    • You may notice that connected sensors might be off by up to 5% before calibration
    • This is because the app has a default "Count Number" setting that needs to be changed for each sensor to improve accuracy
    • For how to access "Count Numbers" see previous section
    • To calibrate, collect the Count numbers at 0% and 100% strain. Record/note the Counts numbers when the sensor is relaxed for 0% and stretch the sensor 50 mm and record the count number for 100%. These two numbers are min and max that needs to be replaced with App's default setting.
    • For calibration highlight the "Connection" Icon with green color
    • Hit the pro icon with letter i
    • Replace the calibration low and calibration high numbers with the numbers collected at 0% and 100% for the specific sensor you are calibrating.


...